2.19 Surgical Stretcher Back Section—Release Handle/Gas Spring Malfunction
Chapter 2: Troubleshooting Procedures
2.19 Surgical Stretcher Back Section—Release Handle/Gas Spring
Malfunction
NOTE:
The PACU extenders can either be removed from the stretcher, in the armboard
position, or in the extender position. If they are in the extender position
(around the articulating head section), then the release handle on the PACU
extenders must be used to raise or lower the back section.
1.
The gas spring releases and locks (every 10° to 15°) throughout its full
range of motion, and the handle engages and disengages approximately
half-way through its full travel. The handle returns to its original position
without binding.
Yes
No
↓
→
Adjust the surgical back section gas spring (refer to procedure
4.16). If this solves the problem, go to “Final Actions” on page
2-11.
2.
When the handle is engaged approximately half-way through its full travel,
the back section rises without assistance.
Yes
No
↓
→
Replace the surgical back section gas spring (refer to procedure
4.16). If this solves the problem, go to “Final Actions” on page
2-11. Otherwise, call Technical Support.
3.
Go to “Final Actions” on page 2-11.
2
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 2 - 31
Service Manual (144386 REV 2)