2.25 Trauma Stretcher Cassette Lift Handle Malfunction
Chapter 2: Troubleshooting Procedures
2.25 Trauma Stretcher Cassette Lift Handle Malfunction
1. The cassette lift cable and spring are attached correctly.
Yes
No
↓
→
Make sure the round vinyl cap is attached to the open end of the
spring. Move the spring so the open end that attaches to the cable
link is towards the base of the stretcher.
2.
The cassette lift mechanism is obstructed.
Yes
No
↓
→
3.
Remove the obstruction from the cassette lift mechanism or the upper
frame cassette support surface. If this solves the problem, go to “Final
Actions” on page 2-11. Otherwise, call Technical Support.
4.
Go to “Final Actions” on page 2-11.