2.11 Procedural or Surgical Stretcher—Knee Section Will Not Raise or Lower
Chapter 2: Troubleshooting Procedures
2.11
Procedural or Surgical Stretcher—Knee Section Will Not Raise
or Lower
1.
The crank handle (A) is attached to the knee screw assembly (B) (see
figure 2-6 on page 2-23).
Yes
No
↓
→
Attach the crank handle (refer to procedure 4.18). If this solves
the problem, go to “Final Actions” on page 2-11. Otherwise,
continue to step 2.
Figure 2-6. Knee Screw Assembly
2
2.
Another part of the stretcher interferes with the knee screw assembly.
Yes
No
↓
→
3.
Remove the interference.
This solves the problem.
Yes
No
↓
→
4.
Go to “Final Actions” on page 2-11.
5.
Replace the knee screw assembly (refer to procedure 4.18).
This solves the problem.
Yes
No
↓
→
Call Technical Support.
6.
Go to “Final Actions” on page 2-11.
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 2 - 23
Service Manual (144386 REV 2)