4.6 Procedural (P8000), Surgical (P8010), Electric (P8020), Trauma (P8040), or OB/GYN (P8050)
Stretcher—Siderail Latch
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.6 Procedural (P8000), Surgical (P8010), Electric (P8020), Trauma
(P8040), or OB/GYN (P8050) Stretcher—Siderail Latch
Tools required:
Ratchet
T30 Torx®
1
head bit
Torque wrench 0-250 in-lb (0 to 28.2 N•m)
Removal
1.
Remove the latch shoulder bolt (D
)
and nut (G) that attach the latch plate
(F) to the end tube (E) (see figure 4-7 on page 4-15).
Figure 4-7. Siderail Latch
4
2.
Remove the lower pivot bolt (C), roller guide (B), latch bushing (A)
,
and
latch plate (F) from the rail assembly. Discard the lower pivot bolt and any
damaged parts.
Replacement
1.
Align the latch plate (F), the roller guide (B), and the latch bushing (A)
with the bolt holes in the rail assembly.
1. Torx® is a registered trademark of Acument Intellectual Properties, LLC.
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 4 - 15