7.10 Push Handles
Chapter 7: Accessories
7.10 Push Handles
The push handles are located at the head end of the stretcher. They increase the
caregiver’s control over the stretcher during transport. The push handles fold
down and out of the way when not in use.
Installation
Tools required:
Ratchet
3/16" hex head bit
Torque wrench 0-250 in-lb (0 to 28.2 N•m)
Phillips head screwdriver
1. Install two screws (E) to secure the push release handle (D) to the right-
hand push handle latch assembly (C) (see figure 7-9 on page 7-15).
Figure 7-9. Push Handle Installation
2.
Install the shoulder screw (B) through the right-hand push handle latch
assembly (C) and into the upper frame.
3.
Tighten the shoulder screw (B) to 100-140 in-lb (11.3-15.8 N·m) of torque.
4.
Put the right push handle tube (A) in position on the upper frame and align
the holes in the right push handle tube (A) with the holes in the bracket on
the upper frame.
7
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 7 - 15
Service Manual (144386 REV 2)