Cleaning and Care
Chapter 6: General Procedures
Spray Wash (Not applicable to Electric (P8020) Stretcher,
OB/GYN (P8050) Stretcher with the optional exam light
installed)
CAUTION:
Do not spray wash the Electric (P8020) Stretcher or the OB/GYN
(P8050) Stretcher with the optional exam light installed. Electronic
components are not protected from fluid ingress. Equipment damage
could occur.
CAUTION:
Do not directly spray the hydraulic cylinders. Equipment damage could
occur.
CAUTION:
Do not directly spray the scale components. Equipment damage could
occur.
CAUTION:
Do not exceed 1750 psi (12066 kPa) during the spray wash. Equipment
damage could occur.
The stretcher can be spray washed as necessary. Use a
maximum
nozzle
pressure of 1750 psi (12066 kPa) at 24" (61 cm).
Do not
use a pencil point
spray. The temperature of a spray wash that is water only should not be more
than180°F (82°C). The temperature of a spray wash that contains detergent or
solvents (
no bleach
) should not be more than 120°F (50°C).
Do not
spray
under the base shroud. After the spray wash, prepare and paint over any
exposed or chipped steel parts or oxidized areas.
6
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 6 - 3
Service Manual (144386 REV 2)