4.11 Electric (P8020) Stretcher—Nurse Control Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
CAUTION:
Use care when you remove the individual nurse control switch wires.
Damage to the terminals could occur if care is not used.
4.
Remove the nurse control unit (D), and remove the individual nurse control
switch wires (C) from the defective switch.
5.
Squeeze the inside grips and pop the switch (E) from the faceplate to
remove any defective nurse control switch (two types).
Replacement
1.
Press the new switch (E) into the opening on the faceplate of the nurse
control unit (D) to replace any defective nurse control switch.
2.
Connect the individual nurse control switch wires (C) (see figure 4-14 on
page 4-29) to the new switch as shown (see figure 4-15 on page 4-30).
Figure 4-15. Nurse Control Switch Wire Diagram
Violet
Red/White
Red/Red
Brown
Blue
Green
Gray
Brown/Brown
Brown/Brown
3.
Install the two screws (B) and the cable clamp (A) to attach the nurse
control unit (D) to the frame (see figure 4-14 on page 4-29).
4.
Plug the stretcher into an applicable power source.
5.
Do the “Function Checks” on page 2-3.
Page 4 - 30
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)