4.9 Electric (P8020) Stretcher—Left or Right Siderail Cable Assembly and Upright Tube
Chapter 4: Removal, Replacement, and Adjustment Procedures
11. Remove the two screws (N) that attach the cover housing (O) under the
patient control switch (R). Remove the cover housing, and disconnect the
connector (see figure 4-12 on page 4-24).
12. Remove the screw (M) and washer (P) that attach the cable clamp (Q) to
the cable assembly (L).
CAUTION:
Do not use a screwdriver to remove the top rail ratchet rivet. The upper
pivot bracket is made of plastic. Damage to the bracket could occur.
13. Drill a hole through the two top rail ratchet rivets (A). Remove the rivets.
14. Remove the upper roller guide (C) and the plastic upper pivot (B) from the
top of the upright tube (D).
15. Remove the lower pivot bolt (J) that attaches the upright tube (D) to the
bottom rail.
16. Remove the cable assembly (L) with upright tube (D), roller guide (I), and
the two wave washers (H) from the siderail assembly.
Figure 4-12. Left or Right Siderail Cable Assembly and Upright Tube
Page 4 - 24
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)