2.22 Electric Stretcher—Knee Section Does Not Raise or Lower Electrically
Chapter 2: Troubleshooting Procedures
2.22 Electric Stretcher—Knee Section Does Not Raise or Lower
Electrically
1.
The stretcher has been connected to an external power supply, and the
power cord is not damaged.
Yes
No
↓
→
Go to “Initial Actions” on page 2-1.
2.
On the nurse control, the
Head ON/OFF
control is in the
ON
position, and
the indicators are on (see figure 2-2 on page 2-9).
Yes
No
↓
→
3.
Push the
Head Up
or
Head Down
control on both siderails (see figure 2-1
on page 2-8). The knee drive operates in both directions for both sides.
Yes
No
↓
→
4. Go to “Final Actions” on page 2-11.
5. The plugs in the control board are all connected correctly.
SHOCK HAZARD:
Failure to unplug the stretcher from its power source could cause
personal injury or equipment damage.
CAUTION:
Failure to wear an antistatic strap when handling electronic components
could cause component damage.
Yes
No
↓
→
Unplug the stretcher from its power source. Put on an antistatic
strap, and then connect the plugs firmly into position. Go to step
1. If this solves the problem, go to “Final Actions” on page 2-11.
Otherwise, call Technical Support.
6. Go to “Final Actions” on page 2-11.
Page 2 - 34
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)