Trauma Stretcher Function Checks
Chapter 2: Troubleshooting Procedures
Trauma Stretcher Function Checks
Cassette Lift
1.
At each side of the stretcher, pull the cassette lift handle, and turn it 180°.
The cassette lift handles function correctly.
Yes
No
↓
→
2.
Go to “Final Actions” on page 2-11.
2
OB/GYN Stretcher Function Checks
Foot Support
1.
Press the release button underneath the foot supports to raise and lower the
foot supports, and turn them in all directions (forward, backward, up, and
down). The foot supports move freely with no evidence of binding along
both axis. The gas spring for each support releases and locks throughout its
full range of motion, and disengages completely. The gas spring returns to
its original position with no evidence of binding.
Yes
No
↓
→
Foot Section
2.
Squeeze the release handle underneath the foot section. The foot section
releases and is easily stored under the sliding patient platform.
Yes
No
↓
→
3.
Go to “Final Actions” on page 2-11.
Final Actions
1.
Complete the preventive maintenance procedures. See “Preventive
Maintenance” on page 6-5.
2.
Make sure all wireforms and components are in position and correctly
attached.
3.
Complete all necessary administrative work.
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 2 - 11
Service Manual (144386 REV 2)