2.18 Procedural Stretcher—Auto Contour™ Feature or BackSaver Fowler® Feature Does Not Raise
or Lower Correctly
Chapter 2: Troubleshooting Procedures
2.18 Procedural Stretcher—Auto Contour™ Feature or BackSaver
Fowler® Feature Does Not Raise or Lower Correctly
1.
Make sure the cables are not bent or damaged. Both cables are in good
condition.
Yes
No
↓
→
Replace the cables as necessary.
2.
Make sure the head-end gas cylinder operates correctly. The gas cylinder
operation is correct.
Yes
No
↓
→
Adjust the cables or replace the head-end gas cylinder as
necessary. If this solves the problem, go to “Final Actions” on
page 2-11. Otherwise, call Technical Support.
3.
Go to “Final Actions” on page 2-11.
Page 2 - 30
Hill-Rom® Transport, Procedural, and Specialty Stretchers