7.20 Temporal Wrist Rest (P262A02)
Chapter 7: Accessories
4.
Put the temporal wrist rest (C) in a satisfactory position, and tighten the
height adjustment knob (E).
NOTE:
The top of the temporal wrist rest can be turned away from the patient for
increased patient access when required.
Removal
1.
Loosen the mounting knob (D) located under the headrest (F).
2.
Remove the temporal wrist rest (C) from the headrest (F).
Page 7 - 34
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)