4.23 OB/GYN (P8050) Stretcher—Upper Frame Catch Shroud
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.23 OB/GYN (P8050) Stretcher—Upper Frame Catch Shroud
Tools required:
Rubber mallet/hammer
Screwdriver
Removal
1.
Remove the two screws (A) that attach the foot section rest blocks (B) to
the upper frame fixed weldment (C) (see figure 4-34 on page 4-69)
Figure 4-34. Upper Frame Catch Shroud
4
2.
Use the rubber mallet/hammer to tap on the underside of the catch shroud
to loosen the roll pins (D) that attach the foot section rest blocks (B) to the
upper frame fixed weldment (C).
NOTE:
The back of the catch shroud is held in position with a Velcro®
1
strip.
3.
Remove the catch shroud (E) from the stretcher.
1. Velcro® is a registered trademark of Velcro Industries, BV (a Dutch corporation).
Hill-Rom® Transport, Procedural, and Specialty Stretchers