4.21 Electric (P8020) Stretcher—CPR Release Cable Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Adjustment
Adjust the cable adjusting nut (I) (see figure 4-3 on page 4-8) on the threaded
stud so the handle on the CPR handle weldment just touches the plastic CPR
handle when the CPR latch (L) (see figure 4-29 on page 4-61) is fully
extended.
Do the “Function Checks” on page 2-3.