4.15 Procedural (P8000), Transport (P8005), Trauma (P8040), or OB/GYN (P8050) Stretcher—Back
Section Gas Spring
Chapter 4: Removal, Replacement, and Adjustment Procedures
CAUTION:
Do not overextend the retaining rings. If the retaining rings are
overextended, use new ones for the installation. Failure to do so could
cause equipment damage.
4.
Remove the retaining ring (F) and the bottom headed pin (E) at the bottom
of the gas spring.
5.
Turn the gas spring clockwise or counterclockwise to adjust.
WARNING:
Do not use retaining rings that are overextended. Personal injury and
equipment damage could occur.
6.
Install the bottom headed pin (E) and the retaining ring (F) on the bottom
of the gas spring.
7.
Make sure the back section release handle (I) operate correctly, and if
necessary, adjust the gas spring again.
8.
Tighten the jam nut (H).
9.
Do the “Function Checks” on page 2-3.