Preventive Maintenance
Chapter 6: General Procedures
Function
Procedure
Leakage current
Unplug the stretcher from its power source.
Connect the stretcher to the safety tester and connect
this device to the power outlet.
Check that the stretcher is correctly supplied by the
test device: control unit LEDs lit (unlock the functions
if necessary).
Measure the leakage current measures without
operating any function.
The value must be less than 100 µA.
Check the AC power cable and power supply unit if
the value is outside of the specifications (0.2 Ohms).
Replace the AC power cable or power supply unit if
necessary.
Assign an inspection sticker to the stretcher.
Control board
mounts
Make sure the board mounts are secure and in place.
OB/GYN Stretchers
Store-away foot sec- Inspect for proper operation and latching after
tion
removal, and storage.
Inspect the foot section latch/release mechanism for
proper operation.
Gliding litter
Inspect for proper slide engagement of the carriage
along the upper frame.
Inspect the foot support release handles for proper
operation.
Verify that the carriage locks in both the exam and
procedural positions.
Integrated fiber
Inspect the power cord for nicks, cuts, and breaks.
optic exam light
Check the gooseneck for proper positioning and stabil
ity.
Check the fiber optic light for proper illumination.
6
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 6 - 11
Service Manual (144386 REV 2)