2.12 Scale Display Does Not Come On
Chapter 2: Troubleshooting Procedures
2.12 Scale Display Does Not Come On
1.
A check of the battery voltage at connector P4 shows the voltage is greater
than 3.0 VDC.
Yes
No
↓
→
Replace the batteries (refer to procedure 4.32). If this solves the
problem, go to “Final Actions” on page 2-11. Otherwise, go to
step 2.
2.
Make a note of the connections for the load beam cables, and then
disconnect the cables from the scale P.C. board. Press and hold the
Weigh
control for 1 second. The
Hands Off
indicator flashes and the display
shows
Err3
.
Yes
No
↓
→
Replace the display P.C. board (refer to procedure 4.34). If this
solves the problem, go to “Final Actions” on page 2-11.
Otherwise, go to step 3.
3.
Connect one load beam cable to its correct connection on the scale P.C.
board. Press and hold the
Weigh
control for 1 second. The
Hands Off
indicator flashes and the display shows a weight or
Err3
. Repeat this step
for each load beam cable.
Yes
No
↓
→
The load beam or one of its cables is damaged. Repair or replace
as necessary (refer to procedure 4.33). If this solves the problem,
go to “Final Actions” on page 2-11. Otherwise, call Technical
Support.
4.
Go to “Final Actions” on page 2-11.
Page 2 - 24
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)
Содержание P8000
Страница 2: ......
Страница 14: ...NOTES Page xii Hill Rom Transport Procedural and Specialty Stretchers Service Manual 144386 REV 2 ...
Страница 441: ...Schematic Wiring Diagram Control Board P N 49603 ...
Страница 442: ...Schematic Wiring Diagram Electric Stretcher P N 60389 ...
Страница 443: ......