4.8 OB/GYN (P8050) Stretcher—Patient Grip Handle
Tools required:
Ratchet
T30 Torx®
1
head bit
1/8" pin punch
Drill
3/16" drill bit
Hammer
Torque wrench 0-250 in-lb (0 to 28.2 N•m)
Removal
1.
Remove the end tube. Refer to procedure 4.7.
2.
For easy access to the patient grip handle, put the end tube (A) at an angle
toward the foot end of the bed (see figure 4-11 on page 4-21).
3.
Remove the damaged patient grip handle (B) from the end tube (A).
Figure 4-11. Patient Grip Handle
Replacement
1.
Gently slide the patient grip handle on to the end tube (A).
1. Torx® is a registered trademark of Acument Intellectual Properties, LLC.
4
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 4 - 21
Service Manual (144386 REV 2)