4.13 Electric Stretcher—Potentiometer
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.13 Electric Stretcher—Potentiometer
Tools required:
Ratchet
head bit
wrench
½" open end wrench
Side cutters
6" straight edge
Removal
1. Raise and secure the head section to get access to the electric box.
SHOCK HAZARD:
Failure to unplug the unit from its power source could cause personal
injury and equipment damage.
2. Unplug the stretcher from its power source.
3.
Remove the two screws (A) and lockwashers (E) that attach the electric
box cover (C) (see figure 4-16 on page 4-31).
4. Remove the electric box cover (C).
5.
Remove the potentiometer cable assembly connector (B) from P1 on the
control board (H) (see figure 4-17 on page 4-32).
6.
Remove the nut (H) that attaches the potentiometer cable assembly (A) to
the electric box (M). The nut will stay with the wiring assembly (see figure
4-18 on page 4-35).
7. Remove the potentiometer cable assembly (A) from the electric box (M).
8.
Cut the cable ties (I) that attach the potentiometer cable assembly (A) along
the left side of the upper frame.
1. Torx® is a registered trademark of Acument Intellectual Properties, LLC.
2. Allen™ is a trademark of Industrial Fasteners, Inc.
Page 4 - 34
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)