4.33 Scale—Load Beam
Chapter 4: Removal, Replacement, and Adjustment Procedures
Replacement
1. Do the removal procedure in reverse order.
NOTE:
The load beam cables are specific to the ports they connect to on the scale P.C.
board. On the yellow sleeve of each load beam cable there are numbers 1
through 4. Put a circle around the port number for the particular cable:
• Left head-end cable—1
• Left foot-end cable—2
• Right foot-end cable—3
• Right head-end cable—4
2. Calibrate the scale. Go to “Calibration” on page 4-107.
Page 4 - 104
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)