Preventive Maintenance
Chapter 6: General Procedures
Function
Procedure
Telescoping calf
supports
Check the tightening/release mechanism for proper
operation. Verify free rotation to make sure that there
is no binding. Verify when the grip is rotated to the
right, the ball joint is tightened. Repair or replace as
required.
Foot support gas
spring
Perform the following:
Make sure the locknuts on the shoulder bolts at the
foot rest weldment interface are installed flush and
fully seated with the mating part.
Patient grip handles Inspect for
cuts, wear, etc. Replace if necessary.
Accessories
Check the proper operation of any accessories
installed on the bed.
Replace any missing or damaged parts.