4.34 Scale—Display or Scale P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
5.
Set the scale enclosure top (B) aside.
6.
Disconnect the battery cable from P4 on the display P.C. board (F).
7.
Make a note of the cable connections, and then disconnect all cables from
the applicable P.C. board (F or G).
8.
Remove the P.C. board (F or G).
Replacement
1.
Do the removal procedure in reverse order.
2.
If the scale P.C. board was replaced, do as follows:
a.
Configure the new board to the original configuration (see table 1 on
page 4-110).
Table 1: Diagnostic Menu
Menu Option
Function
0
Engages Diagnostic mode
1
Exits Diagnostic mode
2
Engages test mode and a continuous display
3
Shows weight in kilograms and pounds
4
Shows weight in kilograms only
5
Shows weight in pounds only
6
Engages calibration mode in pounds
7
Engages calibration mode in kilograms
8
Engages or disengages the
Enable
control (If you hear
one beep, the
Enable
control is disengaged; if you
hear two beeps, the
Enable
control is engaged.
b. Calibrate the scale (see “Calibration” on page 4-107).
Page 4 - 110