Preventive Maintenance
Chapter 6: General Procedures
Function
Procedure
Motor mounts
Inspect for cracks in weldments.
AC power cord
Examine the plug for damage. Make sure the plug is a
one-piece molded plug assembly. If it is not, replace
the plug cord assembly. Replace any plug cord assem
bly that shows any of these:
• Discoloration of the plug molding around the
plug blades; this could occur if the plug blades
have overheated or arced.
• Any signs of cracking; this could occur if the
plug has been bent and straightened to a point
past its useful life.
• Loose fit of the plug blade (the plug blade
moves in the molding); this could occur if the
molding has overheated or the blades have been
bent and straightened to a point past their useful
life.
Replace the power cord, if damaged.
Nurse controls
Test all the bed function switches on the nurse control
panel for proper operation and function.
Check for the head section to raise max 90° using the
nurse control.
Patient controls
Test all the bed function switches on the patient con
(both siderails)
trol panel for proper operation and function.
Check for the head section to raise max 65° using the
patient control.
Back section manual
crank
Inspect for proper operation and lubrication.
Page 6 - 10
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)