4.16 Surgical (P8010) Stretcher—Back Section Gas Spring
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.
Remove the locknut (C) and the shoulder bolt (D) on the bottom of the gas
spring (B).
5.
Turn the gas spring (B) clockwise or counterclockwise in half-turn
increments to adjust.
6.
Loosely install the shoulder bolt (D).
7.
Make sure the release handle operates correctly, and if necessary, adjust the
gas spring for the
right
release handle, the left PACU release handle, and
the right PACU release handle.
8.
When all release handles are operate correctly, install the locknut (C) on
the bottom of the gas spring.
9.
Tighten the jam nut (A).
10. Do the “Function Checks” on page 2-3.