7.7 IV Transporter (P491)
Chapter 7: Accessories
7.7 IV Transporter (P491)
The IV transporter attaches to the base of the stretcher and allows for the
transport of a portable IV pole without the assistance of the caregiver.
The IV transporter can be mounted on either the right or left side of the
stretcher at the head end only.
Installation
Tools required:
Ratchet
¾" socket
Drill
¼" drill bit
7/8" drill bit
Retaining ring removal/installation tool
Torque wrench 0-250 in-lb (0 to 28.2 N•m)
Safety goggles
1.
Decide which side of the stretcher you will mount the transporter.
NOTE:
There are dimples on the inside surface on both sides of the head end of the
shroud. These identify the correct location for the transporter hole.
2.
Lift the shroud, and find the dimple at the head end of the shroud that
corresponds to the location where you want to mount the transporter.
WARNING:
Failure to wear eye protection may cause eye injury.
3.
Put on the safety goggles.
4.
Drill a ¼" (6.4 mm) pilot hole through the center of the dimple.
5.
Use the pilot hole as a guide, and carefully drill a 7/8" (22.2 mm) hole
through the shroud.
6.
Install the rubber grommet (D) into the hole you just drilled in the shroud
(see figure 7-6 on page 7-11).
Page 7 - 10
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)
Содержание P8000
Страница 2: ......
Страница 14: ...NOTES Page xii Hill Rom Transport Procedural and Specialty Stretchers Service Manual 144386 REV 2 ...
Страница 441: ...Schematic Wiring Diagram Control Board P N 49603 ...
Страница 442: ...Schematic Wiring Diagram Electric Stretcher P N 60389 ...
Страница 443: ......