8 Troubleshooting
Hauser
127
E063
Mainte-
nance
Current limit
0/4 mA output 1
Measured value outside specified current range:
Check measured value for plausibility, if
necessary adjust current output assignment
0/4 mA and/or 20 mA.
yes
no
no
E064
Mainte-
nance
Current limit 20 mA
output 1
yes
no
no
E065
Mainte-
nance
Current limit 0/4 mA
output 2
yes
no
no
E066
Mainte-
nance
Current limit 20 mA
output 2
yes
no
no
E067
Mainte-
nance
Reference value exceeded
controller LS 1
Dosing devices defective;
chemical supply empty;
measured value incorrect -> check for plausibility
and function;
incorrect control direction set;
incorrect contact assigned;
incorrect control function assigned
yes
no
no
E068
Mainte-
nance
Reference value exceeded
controller LS 2
yes
no
no
E069
Mainte-
nance
Reference value exceeded
controller LS 3
yes
no
no
E070
Mainte-
nance
Reference value exceeded
controller LS 4
yes
no
no
E071
Mainte-
nance
Reference value exceeded
controller LS 5
yes
no
no
E073
Failure
Temperature 1, table
value undershot
Check temperature value for plausibility.
If necessary, adjust or extend table.
yes
no
no
E074
Failure
Temperature 2, table
value undershot
yes
no
no
E075
Failure
Temperature 1, table
value exceeded
yes
no
no
E076
Failure
Temperature 2, table
value exceeded
yes
no
no
E080
Mainte-
nance
Range for current
output 1 too small
Increase measuring range span for current output
assignment
no
no
no
E081
Mainte-
nance
Range for current
output 2 too small
no
no
no
E090
Function
check
CPG service switch active
Check on CPG if service was actually carried out.
no
no
no
E094
Failure
Incompatible sensor
version
Digital sensor and transmitter are not compatible.
Maybe, Ex version of sensor is used with non-Ex
version of transmitter or vice versa
no
no
no
E100
Function
check
Current simulation active
Check if functions were consciously selected.
no
no
no
E101
Function
check
Service function active
no
no
no
E106
Function
check
Download active
Wait for download to end.
no
no
no
E116
Failure
Download error
Repeat download.
no
no
no
E117
Failure
DAT memory module
data error
Check with other DAT memory module; when
writing to DAT: repeat write process
yes
no
no
E127
Failure
Sensor power fail
Sensor communication present but sensor has to
little current.
Check whether Memosens connection is
correctly inserted and locked.
no
no
no
Error
no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning start
Fact
User
Fact
User
Fact
User
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...