6 Commissioning
Topclean S
114
Hauser
pH calibration
"Calibration with manual buffer"
"Calibration with buffer table" /
"Calibration with automatic buffer recognition"
Manual buffer
: The buffer pH value is entered manually. The display then shows the current
measured value.
Buffer table
: In the calibration menu from page 67 onwards, you can set two buffer solutions or
define them yourself. The selected pH value and buffer type are displayed.
Automatic buffer recognition
: The device automatically recognises the used buffer. Select the buffer
types (e.g. E+H) in the Calibration menu.
CHOICE
(default = bold)
INFO
Calibration with manual
buffer
Note in display
Display of the type of on-site calibration selected in the calibration settings.
025.0 °C
(–20.0 ... +150.0 °C)
Enter temperature
(only when "Calibration with MTC" is selected)
Confirm by pressing
F
.
025.0 °C
(–20.0 ... +150.0 °C)
Enter buffer temperature
(only when "Calibration with MTC" is selected)
Confirm by pressing
F
.
Immerse pH electrode in
buffer 1
Handling instructions
Immerse the electrode in buffer 1.
Confirm by pressing
F
.
Temperature 1: 25.0 °C
07.00
(pH –2.00 ... +16.00)
Enter pH value of buffer 1
Time: 10 s
pH 1:
mV 1:
°C:
MTC
7.00
0
25.0
Checks the stability of the calibration (buffer 1)
Wait until the pH measurement is stable:
Time does not count down,
pH value no longer flashes,
Confirm by pressing
F
.
!
Note!
You can set the criteria for the stability check under "
Ñ
> Setup 1 > Calibration >
Cal. settings".
Invalid calibration value
Note in display:
If an error occurred (e.g. incorrect buffer used), this message is displayed.
Immerse pH electrode in
buffer 2
Handling instructions
Immerse the electrode in buffer 2.
Confirm by pressing
F
.
Temperature 1: 25.0 °C
07.00
(pH –2.00 ... +16.00)
Enter pH value of buffer 2
Time: 10 s
pH 1:
mV 1:
°C:
MTC
7.00
0
25.0
Checks the stability of the calibration (buffer 2)
Wait until the pH measurement is stable:
Time does not count down,
pH value no longer flashes,
Confirm by pressing
F
.
Invalid calibration value
Note in display:
If an error occurred (e.g. incorrect buffer used), this message is displayed.
Zero point
Good
Slope
Good
07.00
59.00
Note in display:
Info on electrode.
Information on the zero point, slope and calibration quality.
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...