Topclean S
6 Commissioning
Hauser
69
pH 1.30
(0.05 ... 2.00 pH)
Entry of pH value zero point deviation for the alarm
function
If the zero point deviation from the reference zero point
exceeds the value entered here, an alarm (error no. 033) can be
triggered (error activation see page 65).
Ex.: The electrode has a zero point of 7.00 pH (for electrodes
with inner buffer of 7 pH). You enter a zero point deviation
value of 0.05 pH. Then, an alarm can be triggered at measured
zero points <6.95 pH or >7.05 pH.
off
on
SCC (Sensor Condition Check)
This function monitors the electrode status or the degree of
electrode ageing. Possible status messages: "Electrode OK",
"Low wear" or "Replace electrode". The electrode status is
updated after each calibration. When the "Replace electrode"
message appears, an error message may be displayed (E040,
E041).
!
Note!
This function is only available for glass electrodes. If you are
using a glass electrode and an ISFET sensor, you can use the
SCC function without restriction. However, the SCC function
only monitors the glass electrode.
Function1/2:
Uis 1/2:
off
on
00.00 pH
(0...16 pH)
Isothermic compensation
Activate the isotherm compensation and insert the
isotherm intersection point (Uis).
Function off
: for Hauser electrodes
Function on
: Only if the isotherm intersection point
≠
zero
point of the electrode. The bigger the difference between
isotherm intersection point and zero point, the bigger the
measured error at temperature fluctuations.
Uis
: Enter the intersection point at which the isotherms of the
electrode meet.
!
Note!
When you activate the isothermic compensation, the electrode
has to be calibrated before measuring.
threshold
length
02 mV
010s
Stability
The calibration is only considered as stable, if the deviation of
the mV value does not exceed the given threshold for the set
period of time (length) during calibration.
This function allows you to adjust calibration accuracy and
timing individually to your process.
!
Note!
If your process requires extremely accurate calibration, reduce
the threshold value and increase the period (length) to make
sure that the pH value does not drift.
Calibration timer:
Cal-Timer:
Warning:
___________
Time:
on
0001h
0001:00
Calibration timer
If no calibration is performed in the set time, an error message
appears (E115).
Cal Timer
: on = activate
Warning
: Enter the time within which a calibration must take
place.
Time
: Displays the remaining time until an error message
appears (count down).
CHOICE
(default = bold)
INFO
Hold
pH
7. 00
[E]
Cal. timer
Next
Select[¯ ®]
Param
Cal timer :
Warning :
off
0001h
----------------------
Time :
----
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...