6 Commissioning
Topclean S
104
Hauser
6.6.18
Diagnosis
To enter the menu, proceed as follows:
⇒
CHOICE
(default = bold)
INFO
Error list
Error log
Operating log
Calibration log
Ext. sensor data (digital sensors with Memosens
technology only)
Service
Error list
: Displays the currently active errors.
(Complete error list with description see page 125)
Error log
: Lists the last 30 signalled errors with date and
time.
Operating log
(service code necessary)
: Lists the last 30
registered operating steps with date and time.
Calibration log
: Lists the last 30 calibrations with date and
time.
You can use the right arrow key to obtain further
details on the calibration.
Ext. sensor data
: Lists the data stored in the digital sensor,
e.g. sensor identification, calibration data, operating hours,
etc.
!
Note!
• Use the arrow keys to scroll through the lists.
• Leave the lists by pressing
F
.
Calibration log
1 Enter spec. buffer
Zero point
Slpe
El. condit.
<date> <time>
Enter spec. buffer
: Used calibration method.
Zero point
: Zero point calculated during calibration.
Slope
: Slope calculated during calibration.
Electrode condition
: State of the electrode.
<date> <time>
: Calibration date and time
If you are using a digital sensor with Memosens technology, the following data are displayed when you press
T
:
SN
Sensor change date
<date> <time>
SN
: Serial number of the calibrated sensor.
Sensor change date
: Date and time when the sensor was
changed.
Ext. sensor data (digital sensors with Memosens technology only):
When you select "Ext. sensor data", the transmitter indicates that the sensor data are read from the sensor. When reading is finished, the display automatically
switches to the next menu point. If it does not switch automatically, you can display former sensor data by pressing
F
or return to measuring mode by
pressing
.
Identification
Calib. data
Comp. temperature
Sensor status
Sensor info
Display of all data stored in the digital sensor
!
Note!
External sensor data can only be displayed for digital
sensors with Memosens technology.
pH
7.00
Diag
Edit
Next
Select
E
Hold
Error list
Error log
Operating log
Calibration log
Validation log
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...