Topclean S
8 Troubleshooting
Hauser
125
8.1.1
Error number list: Trouble-shooting and configuration
In the following error list, you can find a description of all error numbers.
For each error number there is also information on whether the factory setting (= Fact.) of this error
triggers
• an alarm
• an error current
• cleaning
To enter the error list, proceed as follows:
!
Note!
The second column shows the assignment acc. to NAMUR work sheet NA64 (failure, maintenance,
function control). The NAMUR error with the highest priority is shown in the status field of the
display.
⇒
pH
7.00
Diag
Edit
Next
Select
E
Hold
Error list
Error log
Operating log
Calibration log
Validation log
Error
no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning start
Fact
User
Fact
User
Fact
User
E001
Failure
Memory defective
Switch instrument off and on.
If necessary corrective maintenance at factory.
yes
no
–
–
E002
Failure
Data error in EEPROM
yes
no
–
–
E 003
Failure
Invalid configuration
Repeat download.
E004
Failure
Invalid hardware code
The new software cannot recognise the module.
E005
Failure
Unknown CPG code
Topclean S is not recognised.
Topclean S is not compatible to Mycom S
software.
yes
no
–
–
E006
Failure
Transmitter 1 defective
Test with new transmitter
yes
no
–
–
E007
Failure
Transmitter 2 defective
yes
no
–
–
E 008
Failure
SCS message sensor 1
Impedance of pH glass membrane too low: check
pH sensor; replace it, if necessary
For ISFET sensors: leak current > 400 nA.
Replace sensor.
yes
no
no
E 010
Failure
Temperature sensor 1
defective
Check temperature sensor and connections.
ISFET: Check selection in field "Select
temperature sensor".
yes
no
no
E012
Failure
CPG communication
failure
Check Topclean S connecting cable.
yes
no
no
E013
Failure
Assembly has not reached
service position
Check assembly position and feedback signals,
Compressed air present?
Check pneumatic lines to the assembly. Assembly
blocked mechanically?
For non-Ex: External power supply 24 V/220V
(not supply from Mycom) present?
Possibly, microfuse is blown.
yes
no
no
E014
Failure
Assembly has not reached
measuring position
yes
no
no
E017
Failure
Data error in CPC30
EEPROM
Switch instrument off and on.
If necessary corrective maintenance at factory.
yes
no
–
–
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...