8 Troubleshooting
Topclean S
126
Hauser
E024
Failure
CPG programme
interrupted
Control input 87/88 switched: Ext. control
criteria.
yes
no
no
E027
Failure
Compressed air failure
Pressure below permitted minimum
connection not correct, test: termin. D3/D4
pressure > 2 bar, contact closed: 0 V;
pressure < 2 bar, contact open: 3.2 V.
yes
no
no
E029
Failure
Sensor self test error
Error detected during sensor self test. Check
sensor. If necessary, replace it.
yes
no
–
E 030
Failure
SCS fault reference
electrode 1
Reference impedance too high:
Check reference element and, if necessary replace
reference or combination electrode
For ISFET sensor: leak current > 400 nA. Replace
sensor.
yes
no
–
–
E 032
Failure
Outside set slope range for
sensor 1
Sensor aged or defective;
Reference aged, defective or diaphragm blocked;
Buffer solutions too old or contaminated;
PML not in the buffer solutions
yes
no
–
–
E 033
Failure
Outside set zero point for
sensor 1
yes
no
–
–
E 034
Failure
Outside set offset range
for sensor 1
yes
no
–
–
E040
Mainte-
nance
SCC / electrode status of
sensor 1 bad
Check sensor, replace if necessary; maybe clean,
(glass membrane blocked or run dry; diaphragm
blocked).
yes
no
–
E043
Mainte-
nance
Buffer difference channel
1 too small
Wrong buffers used;
Buffer entry incorrect; automatic buffer
recognition defective.
yes
no
–
E044
Mainte-
nance
Meas. value channel 1
unstable
PML missing; sensor too old; sensor sometimes
dry; cable or plug defective.
yes
no
–
E045
Failure
Calibration aborted
Repeat calibration and renew buffer solution.
Replace electrode if necessary.
yes
no
–
E053
Failure
Failure three-point step
controller
reserved
yes
no
no
E054
Mainte-
nance
Dosage time alarm
Dosage time exceeded at total dosage.
Dosage interrupted, dosing agent empty or
process too inconsistent.
yes
no
no
E055
Failure
Display range of main
parameter 1 undershot
Measuring line broken,
sensor in air or air cushion in assembly,
Potential matching missing in symmetrical
measurement, static charging in media with
lowest conductivity
yes
no
no
E057
Failure
Display range of main
parameter 1 exceeded
yes
no
no
E059
Failure
Temperature range 1
undershot
Temperature sensor defective;
Sensor line interrupted or short-circuited;
Wrong sensor type selected
yes
no
no
E061
Failure
Temperature range 1
exceeded
yes
no
no
Error
no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning start
Fact
User
Fact
User
Fact
User
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...