6 Commissioning
Hauser
67
6.6.9
Set up 1 – Calibration
Operating mode pH
To enter the menu, proceed as follows:
⇒
⇒
CHOICE
(default = bold)
INFO
Offset
Manual calibration
Special buffer table
Cal. settings
Calibration timer
Calibration menu selection
Offset
: Entry of a fixed value by which the pH value is
displaced.
Manual calibration
:
Initial settings for the functions of the
CAL key.
Special buffer table
:
Edit tables for special buffer.
Cal. settings
:
General calibration settings
Calibration timer
:
Clock for calibration
Offset:
Act. PV 1/2:
Offset 1/2:
07.00
pH
00.00
pH
Enter offset value for pH value
Act. PV
: Current measured value (primary value) with offset
Offset
: pH value difference
When you enter the measuring mode while an offset is active,
"OFFSET" will be shown on the right top of the display.
Manual calibration:
Enter spec. buffer
Manual buffer
Buffer table
Auto. buffer recognition
Calibration parameters
Sets the calibration type undertaken when the "CAL" key is
pressed:
Enter spec. buffer
: Entry of zero point and sensor slope.
Manual buffer
: During calibration, enter the buffer value.
Buffer table
: If the same buffer values are always used, you
can select this function.
Auto. buffer recognition
: The transmitter Mycom S
automatically recognises the used buffer values.
!
Note!
The automatic buffer recognition only functions if glass
electrodes are connected to the transmitter. In case you are
using an ISFET sensor, please calibrate with a different
calibration function.
DIN 19267
Ingold
E+H
NBS / DIN 19266
Special buffer
Select buffer type
(buffer table, auto. buffer recognition only )
Special buffer
= The special buffer tables defined in the option
"Special buffer table" are used.
!
Note!
You can find the buffer tables for the offered buffers in the
Appendix (see page 170).
PARAM
pH
7.00
Param
Parametrieren
Grundeinstellung
Sonderfunktionen
Handbedienung
Quick
Setup
Edit
Weiter E
pH
7.00
Hold
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next E
pH
7.00
Param
Edit
Next
Set up 1
E
Hold
Relays
Temperature
Alarm
Hold
Calibration
Hold
pH
7. 00
Param
[¯ ]
[E]
¯
Calibration
Offset
Manual calibration
Spec. buffer table
Cal. settings
Next
Calibration timer
Edit
Hold
pH
7. 00
Param
[E]
Offset
Act. PV 1
Offset 1:
Next
Select
07.00 pH
00.00 pH
[¯®]
Hold
pH
7. 00
[E]
Cal. buffer
Enter spec. buffer
Manual buffer
Next
Edit[¯ ]
Param
Buffer table
Auto. buffer recogn.
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...