Liquistation CSF39
Diagnostics and troubleshooting
Hauser
153
11.13 Resetting the measuring device
Menu/Diagnostics
Function
Options
Info
Device restart
Selection
• OK
• ESC
Restart and keep all the settings
Factory default
Selection
• OK
• ESC
Restart with factory settings Settings that have
not been saved are lost.
11.14 Information on operating times
The following information is displayed:
•
Operating hours device
:
Displays the total operating hours of the device in days, hours and minutes
•
Operating hours cooling
(only for the version with a climate control module):
Displays the total operating hours of the compressor in days, hours and minutes
•
Overfill sensor
(for version with vacuum pump):
Number of times a safety switchoff has been caused by LF2
•
Dosing valve
(for version with vacuum pump):
Number of times the dosing valve is actuated;
‐
> corresponds to the number of samples
taken
•
Vacuum pump
(for version with vacuum pump):
Displays the pump operating time in hours and minutes
•
Sample totalizer
(for version with peristaltic pump):
Number of all samples taken and sample errors
•
Pump tube life
(for version with peristaltic pump):
Displays how old the tube is in days, hours and minutes
•
Peristaltic pump
(for version with peristaltic pump):
Displays the pump operating time in hours and minutes
This counter must be reset when a tube is replaced.
•
Filter mats
:
Displays the period of usage in days
•
Operating time photometer
:
Displays the hours of operation in hours
•
Operating time cooling module
:
(optional): Displays the period of usage in days.
With
Reset
set the specific counter reading to zero.
11.15 Status of inputs/outputs
Path:
Display
/
Measurement
The following measured values are listed (read only):
• Binary inputs
Current function state: on or off
• Current inputs
Actual current values of all the current inputs available
• Alarm relay
Current function state: on or off