Operation
Liquistation CSF39
80
Hauser
Settings with the Basic program type with 2 bottles
Sampling mode
=
Time paced CTCV
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Sampling interval
(for version with vacuum pump or
peristaltic pump)
00:01:00 ... 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
00:10:00 HH:MM:SS
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
Set the sampling interval.
Dosing volume
(for version with
vacuum pump or sampling
assembly)
Sampling volume
(for version with
peristaltic pump)
Sampling volume
Factory setting
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling
volume.
Set the sampling volume.
In the version with a vacuum pump or
sampling assembly, the volume is taken
from the setup and can only be modified
there.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.
Multiplier
(for version with vacuum pump or
sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using
the multiplier. For example, if the dosing
volume is set to 200 ml, set the multiplier to
2 for a sampling volume of 400 ml. 2
samples are taken in succession when
sampling.
Bottle change mode
Selection
• Number of samples
• Time
• External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
If is selected:
Bottle change mode
Number of samples
Samples per bottle
1 ... 9999
Factory setting
1
Set the number of samples.
If the bottle is full beforehand based on the
calculated level, the system prevents more
samples being added to the bottle. Such
samples are logged as failed samples in the
program logbook. At the same time, the
diagnostics message "Overfill check" (F353)
is also triggered. Set the number of samples.
Having selected
Bottle change mode
Time
External event
Selection
• No bottle change input
configured
•
Binary input
Sx
Factory setting
No bottle change input
configured
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Time interval
00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
00
‐
01:00 DD
‐
HH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.