Liquistation CSF39
Maintenance
Hauser
161
Proceed as follows to calibrate:
Menu/Calibration active
Function
Options
Info
Sample volume
1-point calibration
Distributor position
Selection
• Front
• Bottle x
• Back
Select the distributor position.
Sample volume
20 to 2000 m
Factory setting
100 ml
Set the sample volume.
Start sampling
Action
The progress of the sampling operation is
displayed.
Check whether the sample volume is correct. Press No to enter the sample volume actually taken, e.g.
110 ml.
Press Yes to repeat the sampling.
2-point calibration
Use 2-point calibration for levels that fluctuate greatly. The second sampling point must be either
higher or lower (height difference of at least 1 m).
Distributor position
Selection
• Front
• Bottle x
• Back
Select the distributor position.
Sample volume
20 to 2000 ml
Factory setting
100 ml
Set the sample volume.
Start 1. sampling
Action
The progress of the sampling operation is
displayed.
Check whether the sample volume is correct. Press No to enter the sample volume actually taken, e.g.
110 ml.
Press Yes to repeat the sampling.
Start 2. sampling
Action
The progress of the sampling operation is
displayed.
Check whether the sample volume is correct. Press No to enter the sample volume actually taken, e.g.
110 ml.
Press Yes to repeat the sampling.
12.3 Replacing the pump tube
L
WARNING
Rotating parts
Minor to medium injury possible.
‣
Take the sampler out of service before opening the peristaltic pump.
‣
Secure the sampler against unintentional start-up whilst you work on the opened hose
pump.