Liquistation CSF39
Operation
Hauser
79
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
External event
Selection
• No bottle change input
configured
•
Binary input
Sx
Factory setting
No bottle change input
configured
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Multiple bottles
0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to
separate bottles.
Start condition
Selection
• Immediate
• Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.
Having selected
Start condition
Immediate
:
Sample at start
Selection
• Yes
• No
Factory setting
Yes
Yes
The first sample is taken when the program
is started.
No
The system waits for the interval to elapse
before the first sample is taken.
Having selected
Start condition
Date/time
:
Start date
01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time
00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Stop condition
Selection
• Program end
• Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Assignment bin. output
Selection
• No binary output
configured
•
Binary output
Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs
Settings for the inputs can be made as
described in the "Inputs" section.