Menu/Setup/Inputs
Function
Options
Info
Totalized flow
The totalized flow is calculated when the program is started if you use a sampling program with
volume, flow-paced sampling or time/flow-paced sampling as the start condition. The samples are
taken based on this value. The current totalizer is used for calculating purposes if the totalized flow is
used as the measured value for an enabling or disabling event.
Current totalized flow
- - -
The totalized flow values are displayed.
Reset totalizer
Selection
• Manual
• Automatic
• At program start
Factory setting
Manual
Manual
Reset the counter manually.
Automatic
The counter is reset automatically at
intervals.
At program start
The counter is reset at program start.
Flow rate
- - -
The current flow rate is displayed.
Having selected
Reset totalizer
Manual
Reset totalized flow
Action
The totalized flow currently calculated is set
to zero when the counter is reset.
Having selected
Reset totalizer
Automatic
Interval
Selection
• Daily
• Weekly
• Monthly
Factory setting
Daily
Daily
If a daily interval is selected, set the
Time
menu item.
Weekly
If a weekly interval is selected, set the
Day
of week
and the
Time
menu item.
Monthly
If a monthly interval is selected, set the
Day
of month
and the
Time
menu item.
Having selected
Input variable
Parameter
Meas. value format
Factory setting
#.#
Specify the number of decimal places.
Parameter name
Free text
Assign a name.
Unit of measure
Free text
Enter the engineering unit.
Lower range value
-20 ... 10000
Factory setting
0
Enter a value for the start of the measuring
range. 0/4 mA is assigned to this value as
per your specifications.
-20 ... 10000
Factory setting
10
Enter a value for the end of the measuring
range. 20 mA is assigned to this value as per
your specifications.
Upper range value
0 ... 60 s
Factory setting
0 s
The damping causes a floating average curve
of the measured values over the time
specified.
Having selected
Input variable
Current
Meas. value format
Factory setting
#.#
Specify the number of decimal places.
Damping
0 ... 60 s
Factory setting
0 s
The damping causes a floating average curve
of the measured values over the time
specified.