Operation
Liquistation CSF39
64
Hauser
Menu/Setup/General settings/
Function
Options
Info
Alarm
0 to 10
Factory setting
7
If there is conductive fouling between
the conductivity 1 and conductivity 2
electrode, a diagnostic message is
displayed when this level of fouling is
reached.
Pump tube life
2)
Control
Selection
• Off
• On
Factory setting
On
Indicates the pump hose has to be
exchanged.
Warning
10 to 50 h
Factory setting
30 h
When the tube has been in operation
for this length of time, a diagnostic
message is displayed to indicate that
the tube should be replaced in time.
Alarm
10 to 50 h
Factory setting
30 h
Totalizer
00-00:00 to 49710-06:28
Factory setting
00-00:00
Operating time of the current pump
hose in days, hours and minutes
Reset
Action
The tube life counter is reset to 0:00
h.
Sample temperature
(optional)
Temp. control
Selection
• Off
• On
Factory setting
On
Switch the temperature control of the
sample compartment on or off here.
Sample temperature
2 to 20 °C
Factory setting
4 °C
Set the sample compartment
temperature.
Cooling control
Selection
• Standard operation
• Quick cool down
Factory setting
Standard operation
The temperature regulator is switched
off for a certain time if quick cool-
down is selected.
Defrosting
The automatic defrosting system is preset at the factory. The following menu items can only be viewed
and changed in the Expert mode.
Mode
Selection
• Off
• On
Factory setting
On
Configuration of the automatic
defrosting function
Interval
Selection
• Hourly
• Daily
• Weekly
• Monthly
Factory setting
Hourly
Select the defrosting interval