Maintenance
Liquistation CSF39
166
Hauser
3. Remove the pump tube.
4. Remove any silicone deposits on the roller and the flexible pump bracket.
5. Make sure the roller turns smoothly and evenly
12.4.3 Sample compartment
The sample compartment has a continuous inner plastic lining.
1. Remove the bottle trays and the distribution pan.
2. Remove the bottles.
3. Remove the distribution arm. See also
4. Spray-clean the sample compartment with a water hose.
You can wash the PE and glass bottles in a dishwasher at 60 °C.
12.4.4 Ventilator and liquefier
1
2
A0013898
70
Cleaning the climate control module
1
Liquefier
2
Ventilator
‣
Clean the liquefier and ventilator with compressed air.
12.4.5 Digital sensors
L
CAUTION
Cleaning not switched off during calibration or maintenance activities
Risk of injury due to medium or cleaning agent!
‣
If a cleaning system is connected, switch it off before removing a sensor from the
medium.
‣
If you wish to check the cleaning function and have therefore not switched off the
cleaning system, wear protective clothing, goggles and gloves or take other appropriate
measures.