Maintenance
Liquistation CSF39
164
Hauser
1
2
6
3
4
5
A0013896
67
Vacuum pump
1
Outflow hose
2
Dosing glass
3
Dosing chamber cover
4
Air hose connection
5
Lock for intake hose
6
Thread adapter nut for intake hose
1. Release the thread adapter nut on the intake hose (item 6).
2. Turn the intake hose at the lock (item 5) to the "open" position and pull the hose
upwards to disconnect it.
3. Release the air hose (item 4) and remove the dosing chamber (item 2) from the front
along with the outflow hose (item 1).
4. Open the bayonet lock (item 3) and open the dosing chamber.
5. Clean these parts (hoses, dosing chamber etc.) with water or soapsuds. Use a bottle
brush if necessary.
You can wash the dosing chamber and dosing chamber cover in a dishwasher at
60 °C.
6. Make sure the dosing pipe is set correctly and set the old value if necessary.
7. Reinstall the cleaned parts in reverse order.
Version with peristaltic pump
Clean the wetted parts as follows:
Содержание Liquistation CSF39
Страница 2: ......
Страница 188: ...www addresses endress com ...