Liquistation CSF39
Technical data
Hauser
179
15.6 Current outputs, active
Span
0 to 23 mA
2.4 to 23 mA for HART communication
Signal characteristic
Linear
Signal on alarm
Adjustable, as per NAMUR Recommendation NE 43
• In measuring range 0 to 20 mA (HART is not available with this measuring range):
Failure current from 0 to 23 mA
• In measuring range 4 to 20 mA: Failure current from 2.4 to 23 mA
• Factory setting for failure current for both measuring ranges: 21.5 mA
Load
Max. 500 Ω
Electrical specification
Output voltage
Max. 24 V
Cable specification
Cable type
Recommended: shielded cable
Cross-section
Recommended: shielded cable