Liquistation CSF39
Operation
Hauser
87
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Bottle change mode
Selection
• Number of samples
• Time
• External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
Having selected
Bottle change mode
Number of samples
Samples per bottle
1 ... 9999
Factory setting
1
Set the number of samples.
Having selected
Bottle change mode
Time
Time interval
00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
00
‐
01:00 DD
‐
HH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Bottle synchronization
Selection
• None
• 1. bottle change time
• 1. Time of
bottle number
Factory setting
None
None
The sampling and bottle change times are
not synchronized.
1. bottle change time
Sampling starts with the first bottle. Set the
synchronization time.
1. Time of bottle number
Each bottle is assigned to a specific fill time.
Set the synchronization time and the
weekday.
Start condition
Selection
• Immediate
• Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.
Having selected
Start condition
Date/time
Start date
01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time
00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Stop condition
Selection
• Program end
• Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Assignment bin. output
Selection
• No binary output
configured
•
Binary output
Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs
Settings for the inputs can be made as
described in the "Inputs" section.