11.6 Overview of diagnostic information
11.6.1 Device-specific, general diagnostic messages
No.
Message
Factory settings
Tests or remedial action
S
D
F
202
Selftest active
F
On
Off
Wait for self-test to be finished
216
Hold active
C
On
Off
Output values and status of the channel are on
hold
241
Firmware failure
F
On
On
Internal device error
1.
Update the software.
2.
Contact the Service Department.
3.
Replace the backplane (Service).
242
Firmware incomp.
F
On
On
243
Firmware failure
F
On
On
261
Electronics module
F
On
On
Electronics module defective
1.
Replace the module.
2.
Contact the Service Department.
262
Module connection
F
On
On
Electronics module not communicating
1.
Check the cable connection , replace it if
necessary.
2.
Check the power supply of the sampling
control module.
3.
Contact the Service Department.
263
Incomp. detected
F
On
On
Wrong kind of electronics module
1.
Replace the module.
2.
Contact the Service Department.
284
Firmware update
M
On
Off
Update completed successfully
285
Update error
F
On
On
Firmware update failed
1.
Repeat.
2.
SD card error → use another card.
3.
Incorrect firmware → repeat with suitable
firmware.
4.
Contact the Service Department.
302
Battery low
M
On
Off
Buffer battery of real time clock is low
The date and time are lost if the power is
interrupted.
‣
Contact the Service Department (battery
replacement).
304
Module data
F
On
On
At least 1 module has incorrect configuration data
1.
Check the system information.
2.
Contact the Service Department.
305
Power consumption
F
On
On
Total power consumption too high
1.
Check installation.
2.
Remove sensors/modules.
306
Software error
F
On
On
Internal firmware error
‣
Contact the Service Department.