Operation
Liquistation CSF39
94
Hauser
10.4 Inputs
As standard, Liquistation CSF39 has:
• Two binary inputs
• Two current inputs
• Galvanically isolated from one another
10.4.1 Binary inputs
The binary inputs are used to control the sampler using external signals.
With the , the power supply of 24 V DC from the terminal block in the connection
compartment of the sampler can be used for floating contacts (see the "Electrical
connection" section).
Menu/Setup/Inputs
Function
Options
Info
Binary input
S:x
Mode
Selection
• Off
• On
Factory setting
Off
Switches the function on or off
Input variable
Selection
• Flow rate
• Rainfall
• External event
• Armature end position
detection
(only for version with
sampling assembly)
Factory setting
Flow rate
• Pulse input for connected flowmeters or
rain gages
• Control of sampling functions via external
signals
Having selected
Input variable
Flow rate
Signal slope
Selection
• Low-High
• High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Unit
Selection
• m³
• l
• cf
• gal
Factory setting
m³
Select the unit.
Meas. value format
Factory setting
#.#
Specify the number of decimal places for the
flow.
1 Impulse =
0 to 1000 m
3
Factory setting
10 m
3
Definition of the pulse value, limits are
calculated depending on the unit
Unit of totalized flow
Current totalized flow
- - -
The totalized flow values are displayed.