Liquistation CSF39
Operation
Hauser
109
Source of data
Measured value
Selection
• Bipolar
(only for current outputs)
• U
• Unipolar-
Mathematical
functions
All the mathematical functions can also be used as a data source and the calculated
value can be used as the measured value.
You can monitor the manipulated variable by assigning the controller manipulated
variable to a limit switch (e. g. configure a dosing time alarm).
Menu/Setup/Additional functions/Limit switches/Limit switch 1 ... 8
Function
Options
Info
Cleaning program
Selection
• None
• Cleaning 1 ... 4
Factory setting
None
Use this function to choose which cleaning
instance should start when the limit switch is
active.
Function
Selection
• Off
• On
Factory setting
Off
Activating/deactivating the limit switch
Operating mode
Selection
• Above limit check
• Below limit check
• In range check
• Out of range check
• Change rate
Factory setting
Above limit check
Type of limit value monitoring:
• Limit value overshoot or undershoot→ 55
• Measured value within or outside a range
Limit value
Settings depend on the
measured value
Operating mode
=
Above limit check
or
Below
limit check
1
2
t
MV
t
1
t
2
t
3
t
4
t
5
1
2
t
MV
t
1
t
2
t
3
t
4
t
5
A
B
A0028523
55
Exceeding (A) and undershooting (B) a limit value (without hysteresis and switch-on delay)
1
Limit value
2
Alarm range
t
1,3,5
No action
t
2,4
An event is generated
• If measured values (MV) are increasing, the relay contact is closed when the switch-on point is exceeded
(
Limit value
+
Hysteresis
) and the start delay (
Start delay
) has elapsed.
• If measured values are decreasing, the relay contact is reset when the switch-off point is undershot (
Limit
value
-
Hysteresis
) and after the drop-out delay (
Switch off delay
).