Liquistation CSF39
Product description
Hauser
11
3.2
Equipment architecture
3.2.1
Slot and port assignment
E
85
85
86
86
41
43
42
87
88
97
98
31
31
32
32
SD
Display
1
2
Sensor
0/4 ... 20 mA
Sensor supply
+
+
–
+
+
–
1
2
Service
PK
GY
PK
GY
GY
1
2
–
BN
WH
GN
YE
BN
WH
GN
YE
87
88
97
98
+
–
A
B
+
–
A
B
–
Alarm
HAR
T
Power
(internal)
24VDC
24V
12,5V
3,3V
A0016633-EN
3
Slot and port assignment of hardware and presentation on the display
The electronics configuration follows a modular concept:
• There are several slots for electronics modules.
• These slots are numbered consecutively in the housing. Slots 0 and 1 are always reserved
for the basic module.
• In addition there are also inputs and outputs for the control module. These slots are
labeled "S".
• Each electronics module has one or more inputs and outputs or relays. Here they are all
collectively known as "ports".
• Ports are consecutively numbered per electronics module and are recognized
automatically by the software.
• Outputs and relays are named according to their function, e.g. "current output", and are
displayed in ascending order with the slot and port numbers.
Example:
"Current output 2:1" shown on the display means: slot 2 (e.g. AOR module) : port 1
(current output 1 of the AOR module)
• Inputs are assigned to measuring channels in the ascending order of "slot:port number"
Example:
"CH1: 1:1" shown on the display means:
Slot 1 (base module) : port 1 (input 1) is channel 1 (CH1) and a conductivity sensor is
connected here.