Liquistation CSF39
Operation
Hauser
115
10.6.3 Cleaning programs, sampling technology
L
CAUTION
Risk of injury due to medium or cleaning agent
Cleaning not switched off during calibration or maintenance activities
‣
If a cleaning system is connected, switch if off before removing a sensor from the
medium.
‣
If you wish to check the cleaning function and have therefore not switched off the
cleaning system, please wear protective clothing, goggles and gloves or take other
appropriate measures.
Dosing chamber rinsing
The sampler must be ordered with TSP modification 71265624 to commission this
function.
The following hardware changes have already been made:
R1 R2 R3 R4
H O
2
4R
A0025513
59
Cleaning valve
Control of cleaning valve via relay R4 for dosing
chamber rinsing.
CSFxx
245
246
CSFxx
291
292
+
-
A0025515
60
Cleaning valve
Binary output S:2 is configured for "End of
sampling". or "Bottle change" and is connected to
binary input S:2 which starts the Chemoclean Plus
function.
Cleaning/rinsing the dosing chamber
1. Select
Menu
/
Setup
/
Inputs
/
Binary input
and configure binary input S:2.