Operation
Liquistation CSF39
96
Hauser
Menu/Setup/Inputs
Function
Options
Info
Reset totalizer
Selection
• Manual
• Automatic
• At program start
Factory setting
Manual
Manual
Reset the counter manually.
Automatic
The counter is reset automatically at
intervals.
At program start
The counter is reset at program start.
Having selected
Reset totalizer
Manual
Reset totalized rainfall
Action
The totalized rainfall currently calculated is
set to zero when the meter is reset
manually.
Having selected
Reset totalizer
Automatic
Interval
Selection
• Daily
• Weekly
• Monthly
Factory setting
Daily
Daily
If a daily interval is selected, set the
Time
.
Weekly
If a weekly interval is selected, set the
Day
of week
and the
Time
.
Monthly
If a monthly interval is selected, set the
Day
of month
and the
Time
.
Time
00:00:00 ... 23:59:59
HH:MM:SS
Factory setting
12:00:00 HH:MM:SS
Input variable
External event
Signal slope
Selection
• Low-High
• High-Low
Factory setting
Low-High
Preselect the level change of the signal.
‐‐
> If
Low
‐
High is selected, the high level brings
about the corresponding setting.
Having selected
Input variable
Armature end position detection
(only for version with sampling assembly)
Position
Selection
• Off
• Front (sampling)
• Back (dosing)
Factory setting
Off
Setting specifying the assembly position
(front or back) to which the end position
sensor is connected.
Binary input assignment view
Overview of the outputs to which this binary
input is linked.
10.4.2 Current inputs
The current input must be assigned an analog signal for the functions described. Active
and passive current inputs are available to connect two-wire or four-wire devices.