Maintenance
Hauser
167
Replacing the sensor while ensuring measuring point availability
If an error occurs or the maintenance schedule stipulates that the sensor has to be
replaced, use a new sensor, or a sensor that has been precalibrated in the laboratory.
• A sensor is calibrated in the laboratory under optimum external conditions, thereby
ensuring better quality of measurement.
• You must perform onsite calibration if you use a sensor that is not precalibrated.
1. Remove the sensor that requires maintenance.
2. Install the new sensor.
The sensor data are automatically accepted by the transmitter. A release code is
not required.
Measurement is resumed.
3. Take the used sensor back to the laboratory.
In the laboratory you can get the sensor ready for reuse while ensuring the
availability of the measuring point.
Prepare the sensor for reuse
1. Clean sensor.
For this purpose, use the cleaning agent specified in the sensor manual.
2. Inspect the sensor for cracks or other damage.
3. If no damage is found, regenerate the sensor. Where necessary, store the sensor in a
regeneration solution (→ sensor manual).
4. Recalibrate the sensor for reuse.
12.5 Replacing the rechargeable batteries
First remove the cover of the power unit to replace the optional rechargeable batteries.
L
WARNING
Device is live
Incorrect connection may result in injury or death
‣
Make sure the device is disconnected from the power source before you remove the
cover of the power unit.
2.
1.
3.
A0012831
71
Removing the cover on the power unit
1. Release the screw with an Allen key (5 mm).
2. Remove the cover of the power unit from the front.
3. When reassembling make sure that the seals are seated correctly.