Liquistation CSF39
Operation
Hauser
77
Menu/Setup/Sampling programs
Function
Options
Info
Setup program
New
A list of all the programs created is
displayed. For this reason, it is often helpful
to add a "B" for Basic in the program name.
Program1, which is supplied with the device, is displayed, as is a list of all the programs already created (Basic,
Standard or Advanced programs). You can either create a new program or select an existing one.
If you select an existing program, you can edit, delete, start or duplicate it. Furthermore, you can also see
whether this program is a Basic, Standard or Advanced program. If you are creating a new program, select the
Basic, Standard or Advanced program type.
Basic
Program name
Free text
Use a distinct name for your sampling
program. The program name can be up to 16
characters long.
Bottle configuration
Choice of all possible bottle
combinations
Options:
• 1 x direct distribution
• 2 x direct distribution
Distribution plate
• 1 x PE direct distribution
• 12 x PE direct
distribution
• 24 x PE direct
distribution
The ordered bottle configuration is preset or
the configuration selected in the setup is
displayed.
Bottle volume
0 to 100000 ml
Factory setting
• 30000 ml
• 20000 ml
Set the bottle volume. The preset value
depends on the bottle configuration
configured. The bottle volume is always 30 l
for individual containers. The preset value
depends on the bottle configuration. The
bottle volume is always 20 l for individual
containers.
Sampling mode
Selection
• Time paced CTCV
• Flow paced VTCV
• Time/flow paced CTVV
• External signal
Factory setting
Time paced CTCV
The following functions depend on the
option selected. These versions are
illustrated individually in the following
section to provide a clearer understanding of
the options.
Time paced CTCV
A constant sampling volume is taken at
steady intervals.
In "Advanced" only:
Time monitoring
(min: 00:01:00; max: 99:59:00)
Flow paced VTCV
A constant sampling volume is taken at
variable intervals.
Time/flow paced CTVV
A sampling volume adjusted to the flow rate
is taken at steady intervals. The sampling
volume is calculated based on the current
flow rate or the average value between two
samples.
External signal
Controlled via binary input.